LabThink OX2/231 Oxygen Permeability Tester

LabThink OX2/231 Oxygen Permeability Tester

LabThink OX2/231 Oxygen Permeability Tester

OX2/231 is based on the equal pressure method, and is professional applicable to the determination of oxygen transmission rate of film and package products, including plastic films, composite films, sheeting, plastic bottles, plastic bags and other packages.

Technical Specifications

Specifications Film Test Package Test (customization available)
Test Range 0.01~1000 cm3/m2·d (standard)
0.1~10,000 cm3/m2·d (optional)
0.0001~10 cm3/pkg·d (standard)
Number of Specimens 1~3
Resolution 0.01 cm3/m2·d 0.0001 cm3/pkg·d
Temperature Range 15°C~55°C ±1.5°C
(Constant Temperature Control Device is optional)
/
Test Gas O2 and air (outside of supply scope)
Test Area 50 cm2 /
Thickness ≤3 mm (customization is available for other thickness) /
Specimen Size 108 mm x 108 mm
Test in 100% O2: Specimen should be smaller than Ф120 mm and lower than 360 mm
Test in the Air: No size limitation
Bottles: the inner diameter should be bigger than Ф8 mm, outer diameter should be smaller than Ф42 mm (Standard)
Bags or Boxes: Supported by accessories
Carrier Gas High purity nitrogen with more than 99.999% concentration
(outside of supply scope)
Gas Supply Pressure ≥ 0.28 MPa
Port Size 1/8 inch copper tubing
Instrument Dimension 670 mm (L) x 410 mm (W) x 310 mm (H) 670 mm (L) x 410 mm (W) x 630 mm (H)
Power Supply 220VAC 50Hz / 120VAC 60Hz
Net Weight 48 kg 50 kg

Principle
The pre-conditioned specimen is mounted between the upper and lower chambers at ambient atmospheric pressure. One chamber contains oxygen and the other chamber is slowly purged by a stream of nitrogen. Due to the concentration difference between the two chambers, oxygen molecules permeate through the specimen into the nitrogen side and are taken to the coulometric sensor where proportional electrical signals are generated. The oxygen transmission rate is then obtained by analyzing and calculating the signals. For package samples, high purity nitrogen flows inside the package, and oxygen flows outside.

Standards
This test instrument conforms to the following standards: ISO 15105-2, GB/T 19789, ASTM D3985, ASTM F2622, ASTM F1307, ASTM F1927, JIS K7126-2, YBB00082003

Basic Applications
Films Including plastic films, plastic composite films, paper-plastic composite films, coextruded films, aluminized films, aluminum foils, aluminum foil composite films and many others
Sheeting Including various sorts of engineering plastics, rubber and building materials, e.g. PP, PVC and PVDC
Packages Including plastic, rubber, paper, paper-plastic composite, glass and metal packages, e.g. Coke bottles, peanut oil packages, Tetra Pak materials, vacuum bags, metal three-piece cans, plastic packages for cosmetic, soft tubes for toothpaste, jelly and yogurt cups

Extended Applications
Package Caps Test seal performance of different package caps
Solar Back-Sheets Including solar back-sheets
Plastic Pipes Including various sorts of pipes, e.g. PPR
Blister Packs Test oxygen transmission rate of the whole blister packs
Contact Lenses Test oxygen transmission rate of contact lenses in usage situation

Fuel Tanks of Cars Plastic fuel tanks are widely used in cars for its light weight, buffering vibration and easy molding characters. But its fuel permeability is the most essential factor, this instrument can be used to test permeability of plastic fuel tanks

Battery Plastic Shell Battery electrolyte is protected by the plastic shell from outside environment. Battery service life is directly dependent on its oxygen permeability. This instrument can be used to test oxgyen transmission rate of battery plastic shell

Red Wine BottlesnTest oxygen transmission rate of red wine bottles Biodegradable Films The oxygen transmission rate is an essential factor that affects biodegradable film properties and further development

Soft Plastic Infusion Bottles Oxygen is the main factor that leads injection to deterioration. It will be very important to lower oxygen content of infusion bottles

Warm Paste Packages Once touched with oxygen before usage, warm pastes will lose efficacy, they should be placed in vacuum packages. So it is necessary to test the oxygen transmission rate of warm pastes packages.

Features
Tests 3 equivalent specimens at one operation and exports test results in average value
2 test modes for both films and packages, and accessories for package test are available for customization
Equipped with constant temperature and humidity control devices (optional parts) to meet different test requirements
Reference film for fast calibration to ensure accurate and universal test data
The instrument is controlled by micro-computer with LCD, menu interface and PVC operating panel, which could conveniently export test data, test results and test curves
Test data could be automatically and safely saved by the function of power failure protection
Micro-printer and standard RS232 port for convenient data output and transfer
Supports LystemTM Lab Data Sharing System for uniform management of test results and test reports

Standard Configurations
Instrument, Micro-printer, Professional Software

Optional Parts
Constant Temperature Control Device, Constant Humidity Control Device, Sealing Accessories for Package Test and Test Hood for Package Test

Note
1. The gas supply port of the instrument is 1/8 inch copper tubing;
2. Customers will need to prepare for gas supply.

We do undertake Servicing and repairing of LabThink OX2/231 Oxygen Permeability Tester labthink Range of products throughout india

Apart from supplying labthink Electrical Testing products in India through dealers network, catering to a variety of customers ranging from Atomic energy, Electricity Boards, Electrical Laboratory, Shopping Complexes, Shopping Mall, Electrical Contractors, Engineering Industries, Turnkey Construction Companies, Railways, Defence Installations, Research Laboratories, Educational Institutes and various Research and Development laboratories of leading national and multinational companies.

We DFT TECH, are labthink Authorized Distributors, Importers, Stockists, Dealers and Suppliers of LabThink OX2/231 Oxygen Permeability Tester Products in Chennai, Bangalore, Hyderabad, Pune, Mumbai, Delhi, Calcutta, Ahmedabad, Sricity, Ambattur, Guindy, Sriperumputhur, Oragadam, Tiruvallur, Trivandrum , Coimbatore, Tiruchy and etc..